Meta Data

(Update 18-Jan-2021 UTC)

General information on the meta-data file
DATA_TYPE none
FORMAT_VERSION 3.1
HANDBOOK_VERSION 1.2
DATE_CREATION 2015-06-13 19:05:16
DATE_UPDATE 2015-06-13 19:05:16
Float characteristics
PLATFORM_NUMBER 5900154
PTT 10609
TRANS_SYSTEM ARGOS
TRAN_SYSTEM_ID 02147
TRANS_FREQUENCY 401650000
POSITIONING_SYSTEM ARGOS
PLATFORM_FAMILY FLOAT
PLATFORM_TYPE PROVOR_MT
PLATFORM_MAKER METOCEAN
FIRMWARE_VERSION n/a
MANUAL_VERSION 20/11/01 ver.1.1
FLOAT_SERIAL_NO MT031
STANDARD_FORMAT_ID 100003
DAC_FORMAT_ID P2
WMO_INST_TYPE 841
PROJECT_NAME J-ARGO
DATA_CENTRE JA
PI_NAME JAMSTEC
ANOMALY none
BATTERY_TYPE Lithium
BATTERY_PACKS none
CONTROLLER_BOARD_TYPE_PRIMARY n/a
CONTROLLER_BOARD_TYPE_SECONDARY none
CONTROLLER_BOARD_SERIAL_NO_PRIMARY n/a
CONTROLLER_BOARD_SERIAL_NO_SECONDARY none
SPECIAL_FEATURES none
FLOAT_OWNER JAMSTEC
OPERATING_INSTITUTION JAMSTEC
CUSTOMISATION none
Float deployment and mission information
LAUNCH_DATE 2002-07-07 02:01:23
LAUNCH_LATITUDE 0.0089
LAUNCH_LONGITUDE -160.005
LAUNCH_QC 1
START_DATE 2002-07-07 02:01:23
START_DATE_QC 9
STARTUP_DATE 2002-07-07 02:01:23
STARTUP_DATE_QC 1
DEPLOYMENT_PLATFORM Mirai
DEPLOYMENT_CRUISE_ID MR02-K01 leg2
DEPLOYMENT_REFERENCE_STATION_ID 014
END_MISSION_DATE 2002-07-07 02:01:23
END_MISSION_STATUS T
Sensor information
SENSOR CTD_PRES
SENSOR_MAKER AMETEK
SENSOR_MODEL AMETEK
SENSOR_SERIAL_NO 2657714170
SENSOR CTD_TEMP
SENSOR_MAKER SBE
SENSOR_MODEL SBE41CP
SENSOR_SERIAL_NO 0192
SENSOR CTD_CNDC
SENSOR_MAKER SBE
SENSOR_MODEL SBE41CP
SENSOR_SERIAL_NO 0192
Float parameter information
PARAMETER PRES
PARAMETER_SEMSOR CTD_PRES
PARAMETER_UNITS decibar
PARAMETER_ACCURACY 2.4
PARAMETER_RESOLUTION 0.1
PARAMETER PSAL
PARAMETER_SEMSOR CTD_CNDC
PARAMETER_UNITS psu
PARAMETER_ACCURACY 0.005
PARAMETER_RESOLUTION 0.001
PARAMETER TEMP
PARAMETER_SEMSOR CTD_TEMP
PARAMETER_UNITS degree_Celsius
PARAMETER_ACCURACY 0.002
PARAMETER_RESOLUTION 0.001
Float calibration information
PREDEPLOYMENT_CALIB_EQUATION y = thermistor output; t = PTHA0 + PTHA1 * y + PTHA2 * y^2; x = pressure output - PTCA0 - PTCA1 * t - PTCA2 * t^2; n = x * PTCB0 / (PTCB0 + PTCB1 * t + PTCB2 * t^2); pressure (psia) = PA0 + PA1 * n + PA2 * n^2;
PREDEPLOYMENT_CALIB_COEFFICIENT PA0=-3.115939e+00, PA1=1.413572e-01, PA2=1.436282e-08, PTHA0=5.655342e+01, PTHA1=-5.467398e-02, PTHA2=1.266482e-05, PTCA0=2.393981e+01, PTCA1=-8.004603e-01, PTCA2=-8.436139e-03, PTCB0=1.004112e+02, PTCB1=8.597902e-03, PTCB2=-1.857809e-04
PREDEPLOYMENT_CALIB_COMMENT Calibrated on 24/Sep/2001
PREDEPLOYMENT_CALIB_EQUATION f = INST FREQ * sqrt(1.0 + WBOTC * t) / 1000.0; Conductivity = (g + h * f^2+ i * f^3 + j * f^4) / (1 + d * t + e * p) Siemens / meter; t = temperature[degree C]; p = pressure[decibars]; d = CTcor; e = CPcor;
PREDEPLOYMENT_CALIB_COEFFICIENT g=-9.906706e-01, h=1.498383e-01, i=-2.166162e-04, j=4.041053e-05, CPcor=-9.570000e-08, CTcor=3.250000e-06, WBOTC=1.810400e-05
PREDEPLOYMENT_CALIB_COMMENT Calibrated on 27/Sep/2001
PREDEPLOYMENT_CALIB_EQUATION Temperature ITS-90 = 1/{a0 + a1[ln(n)] + a2[ln^2(n)] + a3[ln^3(n)]} - 273.15 (degree C); n = instrument output;
PREDEPLOYMENT_CALIB_COEFFICIENT a0=2.792550e-05, a1=2.709253e-04, a2=-2.042466e-06, a3=1.506738e-07
PREDEPLOYMENT_CALIB_COMMENT Calibrated on 27/Sep/2001
Configuration parameters
CONFIG_CycleTime_hours 240
CONFIG_ParkPressure_dbar 2000
CONFIG_ProfilePressure_dbar 2000
CONFIG_Direction_NUMBER 1
CONFIG_TransmissionRepetitionPeriod_seconds 30
CONFIG_TransmissionMinTime_hours 12
CONFIG_AscentTime_hours 6.173
CONFIG_ParkTime_hours 210.716
CONFIG_ParkSamplingPeriod_hours 24
CONFIG_ClockAscentStart_HH 13
CONFIG_FloatReferenceDay_FloatDay 0
CONFIG_MaxCycles_NUMBER 255
CONFIG_GroundingMode_LOGICAL 0
CONFIG_MISSION_NUMBER 0
CONFIG_CycleTime_hours 240
CONFIG_ParkPressure_dbar 2000
CONFIG_ProfilePressure_dbar 2000
CONFIG_Direction_NUMBER 1
CONFIG_TransmissionRepetitionPeriod_seconds 30
CONFIG_TransmissionMinTime_hours 12
CONFIG_AscentTime_hours 6.173
CONFIG_ParkTime_hours 210.716
CONFIG_ParkSamplingPeriod_hours 24
CONFIG_ClockAscentStart_HH 13
CONFIG_FloatReferenceDay_FloatDay 0
CONFIG_MaxCycles_NUMBER 255
CONFIG_GroundingMode_LOGICAL 0
CONFIG_MISSION_COMMENT none